Comparison Of Micro-Computerized Tomography And Cone-Beam Computerized Tomography In The Detection Of Accessory Canals In Primary Molars
Tarih
2015Yazar
Acar, Buket
Kamburoğlu, Kıvanç
Tatar, İlkan
Arıkan, Volkan
Çelik, Hakan Hamdi
Yüksel, Selcen
Özen, Tuncer
Üst veri
Tüm öğe kaydını gösterÖzet
Purpose This study was performed to compare the accuracy of micro-computed tomography (CT) and cone-beam computed tomography (CBCT) in detecting accessory canals in primary molars. Materials and Methods Forty-one extracted human primary first and second molars were embedded in wax blocks and scanned using micro-CT and CBCT. After the images were taken, the samples were processed using a clearing technique and examined under a stereomicroscope in order to establish the gold standard for this study. The specimens were classified into three groups: maxillary molars, mandibular molars with three canals, and mandibular molars with four canals. Differences between the gold standard and the observations made using the imaging methods were calculated using Spearman's rho correlation coefficient test. Results The presence of accessory canals in micro-CT images of maxillary and mandibular root canals showed a statistically significant correlation with the stereomicroscopic images used as a gold standard. No statistically significant correlation was found between the CBCT findings and the stereomicroscopic images. Conclusion Although micro-CT is not suitable for clinical use, it provides more detailed information about minor anatomical structures. However, CBCT is convenient for clinical use but may not be capable of adequately analyzing the internal anatomy of primary teeth.
Bağlantı
https://doi.org/10.5624/isd.2015.45.4.205https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4697004/
http://hdl.handle.net/11655/16462